IRSST - Institut de recherche Robert-Sauvé en santé et en sécurité du travail

Silicon carbide (non fibrous) (Pr)  

French name in the regulation: Silicium, carbure de (non fibreux) (Pr)
CAS : 409-21-2

Exposure Standards in RROHS regulation

Permissible exposure values (PEV)
ppm mg/m3 f/cc Notes
TWAEV - 3 - 1; Pr
STEV - - - -
Ceiling - - - -
Carcinogenicity Sensitizer
Simple asphyxiantPercutaneous
Prohibited useProhibited recirculation
Minimum exposureExplosion hazard
Ototoxic

TWA adjustment for unusual work schedules

Undetermined

CNESST – Toxicological Index Service

Consult REPTOX file:

Analysis:

Weight measurement for inhalable and respirable dusts

Method: IRSST - 424
Sampling device:
Sampling media required:
#IRSSTDescription
924 Plastic IOM, PVC filter, 5 µm, 25 mm

Parameters:
Flow rate: 2,0 L/min
TWA sampling volume: 480 L
STEL sampling volume: -
Principle: Gravimetric measurement (Grav)
Remarks: The minimum reported value (MRV) is 60 µg for respirable dusts and 200 µg for inhalable dusts. The analytical uncertainty (CVa) is 2.3% for respirable dusts and 3.4% for inhalable dusts. Additional information is available in Info-Labo 1991-03 and in Info-Labo 2019-06. The weight measurement method for dusts is by definition non-specific.
IRSST. "Détermination simultanée des poussières de fraction respirable et inhalable dans l'air". Méthode 424. Non publié.