Nickel, total fraction

French name in the regulation: Nickel, fraction total
CAS : 7440-02-0

Not in the regulation.


CNESST – Toxicological Index Service

Consult REPTOX file:

Analysis:

Nickel, composés insolubles (exprimé en Ni)

Method: IRSST - 362
Sampling device:
Sampling media required:
#IRSSTDescription
913 MCE filter, 37 mm, 0,8µm, preweighed

Parameters:
Flow rate: 1,5 L/min
TWA sampling volume: 180 L
STEL sampling volume: -
Remarks: Refer to the information in InfoLABO 2018-11 regarding limitations related to gravimetric measurement obtained through the use of these cassette. For surface wipe sampling, contact Laboratory Customer Service.
Principle: Inductively coupled plasma spectrometry (ICP-MS)
Precision (%): 1,9
Minimum reported value: 2 µg
Remarks: This method allows the analysis of chemical elements present in the sample without distinguishing the different chemical forms. The analytical results are expressed as total nickel.
IRSST, Détermination de métaux dans l'air des lieux de travail par spectrométrie de masse avec plasma à couplage inductif (ICP-MS), Méthodes analytiques, / MA-362-fr, Montréal, IRSST, 2010, 17 pages.

Nickel, composés insolubles (exprimé en Ni)

Method: IRSST - 362
Sampling device:
Sampling media required:
#IRSSTDescription
916 MCE filter, 25 mm, 0.8µm, preweighed

Parameters:
Flow rate: 1,5 L/min
TWA sampling volume: 180 L
STEL sampling volume: -
Remarks: Refer to the information in InfoLABO 2018-11 regarding limitations related to gravimetric measurement obtained through the use of these cassette. For surface wipe sampling, contact Laboratory Customer Service.
Principle: Inductively coupled plasma spectrometry (ICP-MS)
Precision (%): 1,9
Minimum reported value: 2 µg
Remarks: This method allows the analysis of chemical elements present in the sample without distinguishing the different chemical forms. The analytical results are expressed as total nickel.
IRSST, Détermination de métaux dans l'air des lieux de travail par spectrométrie de masse avec plasma à couplage inductif (ICP-MS), Méthodes analytiques, / MA-362-fr, Montréal, IRSST, 2010, 17 pages.

Nickel Solu-Sert

Method: IRSST - 394
Sampling device:
Sampling media required:
#IRSSTDescription
990 25-mm Solu-Sert MCE filter

Parameters:
Flow rate: 1.5 L/min
TWA sampling volume: 180 L
STEL sampling volume: -
Remarks: For surface wipe sampling, contact Laboratory Customer Service.
Principle: Inductively coupled plasma spectrometry (ICP-MS)
Precision (%): 3,3
Minimum reported value: 0,2 µg
Remarks: This method allows the analysis of chemical elements present in the sample without distinguishing the different chemical forms. The analytical results are expressed as total nickel.
IRSST. Détermination de la concentration des métaux dans l’air par ICP-MS – Prélèvement effectué sur filtre encapsulé digérable Solu Sert.

Nickel Solu-Sert

Method: IRSST - 394
Sampling device:
Sampling media required:
#IRSSTDescription
992 37-mm Solu-Sert MCE filter

Parameters:
Flow rate: 1.5 L/min
TWA sampling volume: 180 L
STEL sampling volume: -
Remarks: For surface wipe sampling, contact Laboratory Customer Service.
Principle: Inductively coupled plasma spectrometry (ICP-MS)
Precision (%): 3,3
Minimum reported value: 0,2 µg
Remarks: This method allows the analysis of chemical elements present in the sample without distinguishing the different chemical forms. The analytical results are expressed as total nickel.
IRSST. Détermination de la concentration des métaux dans l’air par ICP-MS – Prélèvement effectué sur filtre encapsulé digérable Solu Sert.
Following the amendment of Appendix 1 of the Quebec Regulation respecting occupational health and safety (RSST, RLRQ, c. S-2.1, r. 13), part of Quebec Occupational Health and Safety Act, which came into effect on March 26, 2020, the TWAEV, STEL and Ceiling values for certain substances are now expressed in ppm. For these substances, the IRSST performed the conversion calculations to express these TWAEV, TWA and Ceiling values in mg/m3, which thus appears in its Sampling Guide for Air Contaminants in the Workplace (Guide d'échantillonnage des contaminants de l'air en milieu de travail ) and the associated Substance Data Sheets. However, it is imperative to refer to the values expressed in the RSST which are legally binding.