N-Nitrosopyrrolidine  

French name in the regulation: N-Nitrosopyrrolidine
CAS : 930-55-2

Not in the regulation.


CNESST – Toxicological Index Service

Consult REPTOX file:

Analysis:

N-Nitrosopyrrolidine (NPYR)

Method: IRSST - 371
Sampling device:
Sampling media required:
#IRSSTDescription
2210 Florisil tube with pre-tube

Parameters:
Flow rate: 1-2 L/min
TWA sampling volume: 1000 L
STEL sampling volume: -
Remarks: The recommended sampling volume is 400 to 1000 litres. The samples can be kept for 4 weeks in the dark at 4°C.The simulltaneous determination of 8 nitrosamines can be done on the same tube.
Principle: Gas chromatography with nitrogen-phosphorus detection (GC-NPD)
Precision (%): 6,0
Minimum reported value: 0,04 µg
IRSST. "Analyse des nitrosamines dans l'air". Méthode 371. Non publiée.